As the outbreak pneumonia that caused by coronavirus COVID-19 has spread across continents, more and more people begin to worry that the epidemic would upgrade into a pandemic. Scientists and doctors are working together internationally to study further about this new coronavirus and trying to develop the vaccine as soon as possible.
For laboratory diagnosis, centrifuge is one of the essential equipments in laboratory nucleic acid testing of coronavirus COVID-19. As a manufacturer and enterprise of lab centrifuge, we have the responsibility to give our efforts to fight against this disease. Currently we have 3 models suitable for the clinical and diagnostic laboratory.
Model 1: TGL-20MB
Hitro ohlajena centrifuga
Max. Speed: 20000r/min
Max. RCF: 27800xg
Max. Capacity: 4x100ml
Temperature range: -20oC to 40 oC,
Accuracy: ±2 oC
Timer range: 1min~99min59sec
Motor: Converter motor
Hrup: <55db
Screen: LCD color screen
Acceleration / Deceleration rates: 1--10
Power: AC220V, 50/60Hz, 18A
Neto teža: 70kg
Dimenzije: 620x500x350mm (DxŠxV)
rotor:
Angle Rotor 24x1.5ml, 16000rpm, 23800xg
With aerosol-tight lid
Model 2: XZ-20T
Centrifuga visoke hitrosti
Max. Speed: 20000r/min
Max. RCF: 27800xg
Max. Capacity: 4x100ml
Timer range: 1min~99min59sec
Motor: Converter motor
Hrup: <55db
Screen: LCD color screen
Acceleration / Deceleration rates: 1--10
Power: AC220V, 50/60Hz, 5A
Neto teža: 27kg
Dimenzije: 390x300x320mm (DxŠxV)

rotor:
Angle Rotor 24x1.5ml, 16000rpm, 23800xg
With aerosol-tight lid
Model 3: TD5B
Centrifuga za nizko hitrost
Max. Speed: 5000r/min
Max. RCF: 4760xg
Max. Capacity: 4x250ml
Timer range: 1min~99min59sec
Motor: Converter motor
Hrup: <55db
Screen: LCD color screen
Acceleration / Deceleration rates: 1--10
Power: AC220V, 50/60Hz, 5A
Neto teža: 35kg
Dimenzije: 570x460x360mm (DxŠxV)
rotor:
Swing Rotor 48x 5ml, 4000rpm, 2980xg
including a (stainless steal) rotor arm and 4 (aluminum alloy) rectangular buckets
For blood collection tubes (vacutainers) 5ml(13x100mm)
With aerosol-tight lid
Swing Rotor 48x 2ml, 4000rpm, 2625xg
including a (stainless steal) rotor arm and 4 (aluminum alloy) rectangular buckets
For blood collection tubes (vacutainers) 2ml(13x75mm)
With aerosol-tight lid
The above 3 models and rotors are more needed than often due to the increasingly huge needs from laboratory diagnosis on coronavirus COVID-19. Xiangzhi company is trying the best to ensure the production and supply for these models. And on the order hand we consider the biosafety as prior thing for lab workers, we highly appreciate the advices and suggestions of improving our centrifuge designs that’s able to protect lab workers better.
Lastly, please be aware of the following points when handling with hazard hazardous Material in laboratory:
Working in clinical diagnostic laboratories usually means working with potentially infectious samples like blood or other bodily fluids. But handling infectious microorganisms or harmful chemicals is quite common in research laboratories as well. To ensure the safety of laboratory personnel and to prevent laboratory acquired infections (LAIs) or other health hazards, reasonable precautions must be taken throughout the whole workflow.
Centrifuga je eden od virov aerosolov. Širok razpon dejavnosti – vključno s polnjenjem centrifugalnih epruvet, odstranjevanjem pokrovčkov ali pokrovov iz epruvet po centrifugiranju ter odstranjevanjem supernatantne tekočine in nato resuspendiranjem peletov – lahko vodi do sproščanja aerosolov v laboratorijsko okolje.
Thus, aerosol-tight lid or biocontainment cover is essential for centrifuging hazardous samples, such as blood collection tubes (vacutainers)
Aerosol-tight lids do not prevent formation of aerosols during centrifugation; rather, they ensure aerosols cannot leak from the closed system.
Če se cev zlomi ali pušča, centrifuge ne odpirajte vsaj 30 minut po teku. Ker tega ni mogoče vedno zaznati, preden odprete vedra ali rotor (nenadno neravnovesje je lahko prvi znak zloma cevi), priporočamo, da ves čas počakate vsaj 10 minut, preden odprete posode.
Prav tako morate nalagati in raztovarjati vedra ali rotor v omaro za biološko varnost (zlasti v virologiji in mikobakteriologiji), da zmanjšate tveganje za uhajanje aerosolov.